Mass axis resampling maps every spectrum in a dataset onto one shared m/z axis. It is enabled by default, and by default every parameter is chosen automatically from the acquisition metadata.

This page explains what that automatic choice is, how it is made, and how to override it. For the list of flags alone, see the CLI Reference.


Why resampling exists

An imzML file comes in one of two storage modes:

Mode m/z arrays Consequence
continuous one shared array for the whole dataset pixels are already comparable bin-for-bin
processed a separate array per spectrum pixel 1's column 500 and pixel 2's column 500 are different masses

For processed data -- which includes most centroided vendor exports -- you cannot build a pixel x m/z matrix at all without first agreeing on a common axis. Thyra reports this as needs_resampling, and it is the reason resampling defaults to on rather than off.

Even for continuous data, resampling is often still what you want: it lets you impose a physically sensible bin spacing, restrict the mass range, or match the axis convention of another tool.

To keep the original per-spectrum axes untouched, disable it:

thyra input.imzML output.zarr --no-resample

Note that with --no-resample on processed data, each pixel keeps its own m/z values, and most downstream tools will not be able to treat the table as a matrix of comparable features.


How the automatic choice is made

Three things are decided: the method (how intensity moves onto the new axis), the axis type (how bin widths scale with m/z), and the bin count.

  acquisition metadata
  DataCharacteristics        storage mode, spectrum type, instrument
          │                  name/type/manufacturer, peak density
  InstrumentDetectorChain    first matching detector wins
          ├──────────────► resampling method   (nearest_neighbor | tic_preserving)
          ├──────────────► axis type           (constant | *_tof | orbitrap | fticr)
          └──────────────► bin count           from a target width at a reference m/z

What Thyra looks at

DataCharacteristics collects, from the metadata:

  • whether there is a shared mass axis (continuous) or not (processed)
  • spectrum type -- centroid spectrum (MS:1000127) or profile spectrum (MS:1000128)
  • instrument name, instrument type, and manufacturer
  • average peaks per spectrum; above 5000 the data is recorded as high-density profile
  • format flags for Rapiflex and timsTOF

Spectrum type is whatever the file declares -- MS:1000127 or MS:1000128 -- read from the file description first and then from the rest of the document. Only when a file declares neither does Thyra fall back to guessing centroid from processed storage mode, and it says so in a warning when it does. Processed and centroid are independent: processed means each spectrum carries its own m/z array, which says nothing about whether its peaks are centroided.

Peak density is reported but does not select a method or an axis type. It describes how finely the spectra were sampled, not what acquired them, and a dense profile spectrum can come from a MALDI-TOF, a TOF-SIMS, or an Orbitrap run in profile mode. SCiLS Lab does not guess modality either -- its importer takes it as an argument (--project TIMSTOF|TOF|FT).

The timsTOF flag is a case-insensitive substring match on the instrument name, because Bruker names the family many ways (timsTOF fleX MALDI-2, timsTOF Pro 2, timsTOF SCP, and so on).

Which detector wins

Detectors are tried in a fixed priority order and the first match wins: timsTOF, Rapiflex, FT-ICR, Orbitrap, PHI, generic centroid, then a catch-all default. This table is the actual observed behaviour of that chain:

Metadata Detector Method Axis type
timsTOF, centroid timsTOF nearest_neighbor reflector_tof
timsTOF, profile high-density timsTOF nearest_neighbor reflector_tof
Rapiflex, profile Rapiflex MALDI-TOF tic_preserving constant
Bruker MALDI-TOF Rapiflex MALDI-TOF tic_preserving constant
PHI SmartSoft-TOF .raw PHI SmartSoft-TOF (ToF-SIMS) nearest_neighbor linear_tof
unknown vendor, profile (any density) Unknown (default) nearest_neighbor constant
unknown, centroid ImzML Centroid nearest_neighbor reflector_tof
no usable metadata Unknown (default) nearest_neighbor constant

Why PHI needs its own row

Without it PHI reaches the catch-all default and is reported as constant. Thyra's own answer would still be nearest_neighbor, so nothing breaks inside Thyra -- but preview_msi hands that axis type to callers, and a caller that maps constant to profile-MALDI conventions arrives at tic_preserving onto an equidistant axis. That combination is destructive here: a PHI pixel holds a median of 44 measured points across m/z 0.5--1850, and interpolating between them fabricates intensity in every bin in the gaps, with the TIC rescale hiding it behind a total that still balances. The detector reports the law the data actually follows -- PhiMassAxis steps at a constant flight time, so spacing goes as sqrt(m/z) -- which is linear_tof.

tic_preserving is gated on the source and target axis laws matching

A detector may only ask for tic_preserving if it knows the spacing law of the grid the spectra arrive on, and that law is the one it is asking the target axis to use. Only the Rapiflex route qualifies: RapiflexReader lays every spectrum out with np.linspace, so the source is uniform in m/z, and the axis it requests is constant -- the same law.

Anything else is refused and gets nearest_neighbor instead, with a log line saying so. This is the rule SCiLS Lab applies: TIC-preserving resampling "if all axis types are identical", otherwise interpolation (2026b User Guide, p.80). It is also exactly when Thyra's interpolate-then-rescale operator is exact -- see the !!! danger box below for what it costs off the diagonal.

The Orbitrap and FT-ICR detectors cannot currently fire

Both match on an instrument_type of "Orbitrap" or "FT-ICR", but no reader populates that key: the only place it is set is the Rapiflex reader, which hardcodes "MALDI-TOF". Neither the imzML nor the Bruker .d metadata extractor emits it at all.

Still true: reading the declared MS:1000127/MS:1000128 accession fixed how spectrum representation is detected, not how instrument_type is populated. Nothing writes "Orbitrap" or "FT-ICR" into it, so these two detectors remain unreachable. What did change is which row such data lands on: a profile-mode Orbitrap file now correctly reports profile, so it falls to the default row (constant) rather than being misread as centroid and given reflector_tof.

So Orbitrap and FT-ICR data does not get the orbitrap or fticr axis automatically -- it falls through to the generic centroid or default row above and gets reflector_tof or constant. Set --mass-axis-type orbitrap or --mass-axis-type fticr explicitly if you want the matching physics.

tic_preserving is selected for profile MALDI-TOF, not for high resolution

It is easy to assume the high-resolution analysers get the more elaborate method. They do not. tic_preserving is chosen for profile data, where a peak is spread over many points and rebinning would otherwise change the total ion count. Orbitrap and FT-ICR data is normally centroided, so it gets nearest_neighbor, which is the correct choice for discrete peaks. If you have profile Orbitrap or FT-ICR data, set --resample-method tic_preserving yourself.

Do not combine tic_preserving with a non-uniform axis type

tic_preserving interpolates onto the target axis and then applies a single scaling factor to the whole spectrum. A single factor cannot account for bin widths that vary across the mass range, so pairing it with linear_tof, reflector_tof, orbitrap or fticr suppresses high-m/z ions relative to low-m/z ones. Measured across 300-1100 m/z, two ions of equal abundance come back with their ratio distorted by 1.9x on linear_tof, 3.7x on reflector_tof, 7.0x on orbitrap and 13.4x on fticr.

Auto-selection cannot produce these pairings. tic_preserving is only ever chosen alongside constant, and only on a source grid that is itself uniform in m/z, which is what makes it exact; the detector chain enforces that. You have to ask for the combination with two explicit flags, and Thyra takes you at your word.

If you want a non-uniform axis, use nearest_neighbor, which moves each peak into a single bin and is unaffected by bin width.

The chosen detector, method, and axis type are all logged:

INFO - Detected instrument type: timsTOF
INFO - Selected resampling method: NEAREST_NEIGHBOR
INFO - Selected axis type: REFLECTOR_TOF

Methods

nearest_neighbor

Each target bin takes the intensity of the nearest original m/z value. Peaks stay sharp and stay put, and no intensity is invented between them. This is the right choice for centroid data, where a peak is a single discrete mass and smearing it across neighbouring bins would be wrong.

Consequence worth knowing: on a target axis finer than the source spacing, most bins receive nothing. A dataset resampled from 4,000 source points onto 190,000 bins has exactly 4,000 populated bins per spectrum, and the rest are zero. This is why you should extract ion images by summing over an m/z window rather than picking the single nearest bin -- see the tutorial.

tic_preserving

Linear interpolation onto the target axis, followed by rescaling so the spectrum's total ion current matches the original. This is the right choice for profile data: a peak spans many points, interpolation reconstructs its shape on the new grid, and the rescaling step stops rebinning from quietly changing quantitation.

Use it whenever the total ion count per pixel has to stay comparable before and after conversion.

The total that is preserved is the share of the spectrum inside the axis range. If you crop with --resample-min-mz or --resample-max-mz, intensity outside the window is dropped rather than redistributed over the bins you kept, so the per-pixel total falls by whatever you cropped out -- a cropped window should not claim ions from the parts that were cut away. That share is measured by area, so a window narrower than the spacing between source points still keeps a proportionate amount instead of collapsing to zero.

With the default axis, which spans the dataset's own mass range, nothing is dropped and the total is preserved exactly.

Gaps in the source m/z values

Interpolation has no notion of a gap. Between the last source point before an empty stretch and the first one after it, np.interp draws a straight line, and every target bin in between gets an intensity that was never measured. On sparse or thresholded m/z arrays that is most of the axis: forcing tic_preserving onto bellini.imzML -- 2,222 points per spectrum, median source spacing 0.0072 Da, largest gap 1,269 Da -- puts 80.6% of the output total ion current more than 0.5 Da from any measured point.

--resample-gap-tolerance sets how far a target bin may sit from the nearest source m/z before its interpolated value is discarded instead of trusted. It is the same parameter Cardinal calls tolerance and matter calls tol.

thyra data.imzML out.zarr \
  --resample-method tic_preserving \
  --resample-gap-tolerance 0.1

Masking happens before the TIC rescale, so the intensity returns to the bins that do have measurements behind them rather than being deleted; the per-pixel total is still preserved.

Sizing it: on a source grid of uniform step s, no target bin is ever more than s/2 from a source point, so any tolerance above half the widest gap in your source m/z values changes nothing. Continuous profile data is dense and evenly sampled, which is why the default is no limit -- there is nothing to refuse. Set it when your m/z arrays are sparse or peak-picked and you are asking for tic_preserving anyway.

nearest_neighbor needs no such parameter: it only ever fills a bin some peak was snapped into.


Axis types

The axis type sets how bin width grows with m/z. Each corresponds to the physics of a mass analyser, so that bins track the instrument's real resolving power instead of over-sampling the low end and under-sampling the high end.

Axis type In SCiLS Lab Bin width scales as Rationale
constant Constant (equidistant) constant Da Equidistant bins
linear_tof Axial TOF sqrt(m/z) Linear TOF: flight time t ∝ sqrt(m/z), so equal time bins give sqrt(m/z) mass bins
reflector_tof Orthogonal TOF m/z Constant relative resolution R = m/Δm; bins are uniform in ln(m/z)
orbitrap Orbitrap m/z^1.5 Orbitrap frequency f ∝ 1/sqrt(m/z), so equal frequency bins give m/z^1.5 mass bins
fticr MRMS (Fourier-transform) m/z^2 Cyclotron frequency f ∝ 1/(m/z), so equal frequency bins give m/z^2 mass bins

The five types, and their scaling laws, are the ones SCiLS Lab offers for the common mass axis (2026b User Guide, p.75). Thyra's names are SCiLS's older ones: linear_tof and reflector_tof were "Linear TOF" and "Reflector TOF" in earlier SCiLS versions and are now Axial TOF and Orthogonal TOF, and the FT-ICR type is now called MRMS. The laws are unchanged; only the labels moved.

reflector_tof is the most broadly useful of these: constant relative resolution means constant relative mass accuracy across the whole range, which is what most MS workflows assume.


Bin count

You can set the bin count directly, or specify a target bin width at a reference m/z and let Thyra derive the count from the axis physics. The two are mutually exclusive.

When you specify neither, these defaults apply:

Axis type Default width At reference m/z
linear_tof 17 mDa 300
everything else 5 mDa 1000

The 17 mDa / m/z 300 pairing for linear_tof was chosen to be close to the axis SCiLS Lab produces for FlexImaging data. Treat it as a working default rather than a reproduction: the figure does not appear in the SCiLS Lab 2026b User Guide, and no version that states it has been identified, so nothing here is a claim of exact equivalence. The code comment implementing it says "SCiLS-like ~17 mDa", which is the accurate description. If you need to match a particular SCiLS project, set --resample-width-at-mz and --resample-reference-mz from that project's own axis.

The count is then derived per axis type:

Axis type Bin count
reflector_tof ln(max/min) * (ref_mz / width)
linear_tof (2/k) * (sqrt(max) - sqrt(min)), k = width / sqrt(ref_mz)
orbitrap 2 * (1/sqrt(min) - 1/sqrt(max)) * (ref_mz^1.5 / width)
fticr (1/min - 1/max) * (ref_mz^2 / width)
constant (max - min) / width

Each is the integral of 1 / width(m) across the mass range, so the bin width the axis actually realizes at your reference m/z is the width you asked for. The result is floored at 100 bins.

A worked example, from the tutorial's synthetic dataset -- 250-1200 Da, constant axis, default 5 mDa width:

(1200 - 250) / 0.005 = 190000 bins

which is what the log reports:

INFO - Calculating bins for 5.0 mDa width at m/z 1000.0
INFO - Calculated 190000 bins for constant axis type
INFO - Building resampled mass axis: 250.00 - 1200.00 m/z, 190000 bins

190,000 bins from 4,000 source points sounds extravagant, and on the nearest_neighbor path -- which is what auto picks for this dataset -- it very nearly is free. Each source point lands in exactly one bin, so the non-zero count does not depend on the bin width at all and only the index arrays lengthen: the tutorial store comes out at roughly 31 MB, and going from 4,000 bins to 190,000 costs about +0.8% on disk.

That only holds for nearest_neighbor

tic_preserving interpolates, so it populates essentially every bin. The matrix comes out 99.998% dense and the store becomes n_pixels x target_bins x ~7.5 bytes -- independent of how sparse the source was. Measured on this same 4,000-point source: at 190,000 bins it costs 1.44 MB per pixel, so 400 pixels take 574 MB against 6.93 MB at 4,000 bins, an 82.9x blowup, and the 1,728-pixel tutorial dataset would take roughly 2.5 GB.

On that path the default bin count is a decision rather than a detail.

Pass --resample-bins or --no-resample if you would rather not upsample.

The 10 million bin cap

--no-resample on a processed-mode imzML does not build a physics axis at all. It builds the raw axis: every distinct m/z value in the whole dataset, one column each. When the peak lists genuinely share m/z values that is small. When they do not -- which is the normal case for centroided peak picking, where two pixels almost never report a peak at the same double -- it grows to roughly one column per peak in the entire dataset, and there is no natural stopping point.

Thyra gives up once that axis passes 10 million unique m/z values, and says what to do instead:

ValueError: Common mass axis exceeded 10,000,000 unique m/z values after
12,400 of 918,855 spectra (10,004,112 so far). The peak lists in this dataset
do not share m/z values, so a raw axis grows to roughly one column per peak,
which is not usable downstream. Convert with resampling instead (it is the
default; --no-resample disables it), or raise max_mass_axis_length.

10 million is SCiLS Lab's own limit on the same quantity: "Data sets in SCiLS Lab are limited to a maximum of 10 million bins on the common mass axis" (2026b User Guide, p.76).

The cap is Python-API only, through reader_options, and applies to the imzML reader:

from thyra import convert_msi

convert_msi(
    "data.imzML",
    "out.zarr",
    reader_options={"max_mass_axis_length": 50_000_000},  # or None for no cap
)

It has no effect when resampling is on, which is the default: a resampled axis has the bin count you asked for.


Overriding the automatic choice

Every part is independently overridable; anything you leave alone stays automatic.

# Method and axis type. nearest_neighbor is the pairing to use with a
# non-uniform axis -- see the warning under "Selection" above.
thyra input.imzML output.zarr \
    --resample-method nearest_neighbor \
    --mass-axis-type orbitrap

# Fixed bin count
thyra input.imzML output.zarr --resample-bins 50000

# Target width at a reference m/z instead of a bin count
thyra input.imzML output.zarr \
    --resample-width-at-mz 0.01 \
    --resample-reference-mz 500

# Restrict the mass range
thyra input.imzML output.zarr \
    --resample-min-mz 400 --resample-max-mz 1000

# Off entirely
thyra input.imzML output.zarr --no-resample
from thyra import convert_msi

convert_msi(
    "input.imzML",
    "output.zarr",
    resampling_config={
        "method": "nearest_neighbor",
        "axis_type": "orbitrap",
        "width_at_mz": 0.01,
        "reference_mz": 500.0,
    },
)

The Python API does not resample by default

convert_msi() called without resampling_config keeps the original mass axis, whereas the thyra command-line tool resamples by default. Pass a resampling_config explicitly if you want the CLI's behaviour from Python.


Checking what was used

The decision is recorded in the output store, so a converted dataset is self-describing:

import spatialdata as sd

sdata = sd.read_zarr("output.zarr")
table = sdata.tables["msi_dataset_z0"]

mz = table.var["mz"].values
print(f"{len(mz)} bins, {mz.min():.2f} -- {mz.max():.2f} m/z")

# Bin width across the range tells you the axis type in practice:
# flat -> constant, growing linearly -> reflector_tof, and so on.
import numpy as np
d = np.diff(mz)
print(f"bin width: {d.min()*1000:.3f} -- {d.max()*1000:.3f} mDa")

For the tutorial dataset this prints a flat 5.000 mDa across the range, confirming a constant axis at the default width.


Practical guidance

  • Leave it alone unless you have a reason. The detected settings are right for the common Bruker and imzML cases.
  • Disable it (--no-resample) when you want the untouched vendor axis -- your own peak picking, centroiding, or calibration downstream.
  • Set tic_preserving if quantitation matters and your data is profile, especially profile data from a high-resolution analyser, which the automatic choice will not pick it for.
  • Set --resample-bins any time you need to match another tool's axis exactly.
  • Narrow the mass range (--resample-min-mz / --resample-max-mz) before increasing bin counts; it is usually the cheaper way to get the resolution you need where you need it.