Resampling¶
Mass axis resampling maps every spectrum in a dataset onto one shared m/z axis. It is enabled by default, and by default every parameter is chosen automatically from the acquisition metadata.
This page explains what that automatic choice is, how it is made, and how to override it. For the list of flags alone, see the CLI Reference.
Why resampling exists¶
An imzML file comes in one of two storage modes:
| Mode | m/z arrays | Consequence |
|---|---|---|
| continuous | one shared array for the whole dataset | pixels are already comparable bin-for-bin |
| processed | a separate array per spectrum | pixel 1's column 500 and pixel 2's column 500 are different masses |
For processed data -- which includes most centroided vendor exports -- you
cannot build a pixel x m/z matrix at all without first agreeing on a common
axis. Thyra reports this as needs_resampling, and it is the reason resampling
defaults to on rather than off.
Even for continuous data, resampling is often still what you want: it lets you impose a physically sensible bin spacing, restrict the mass range, or match the axis convention of another tool.
To keep the original per-spectrum axes untouched, disable it:
Note that with --no-resample on processed data, each pixel keeps its own m/z
values, and most downstream tools will not be able to treat the table as a
matrix of comparable features.
How the automatic choice is made¶
Three things are decided: the method (how intensity moves onto the new axis), the axis type (how bin widths scale with m/z), and the bin count.
acquisition metadata
│
▼
DataCharacteristics storage mode, spectrum type, instrument
│ name/type/manufacturer, peak density
▼
InstrumentDetectorChain first matching detector wins
│
├──────────────► resampling method (nearest_neighbor | tic_preserving)
├──────────────► axis type (constant | *_tof | orbitrap | fticr)
└──────────────► bin count from a target width at a reference m/z
What Thyra looks at¶
DataCharacteristics collects, from the metadata:
- whether there is a shared mass axis (continuous) or not (processed)
- spectrum type --
centroid spectrum(MS:1000127) orprofile spectrum(MS:1000128) - instrument name, instrument type, and manufacturer
- average peaks per spectrum; above 5000 the data is recorded as high-density profile
- format flags for Rapiflex and timsTOF
Spectrum type is whatever the file declares -- MS:1000127 or
MS:1000128 -- read from the file description first and then from the rest of
the document. Only when a file declares neither does Thyra fall back to
guessing centroid from processed storage mode, and it says so in a warning
when it does. Processed and centroid are independent: processed means each
spectrum carries its own m/z array, which says nothing about whether its peaks
are centroided.
Peak density is reported but does not select a method or an axis type. It
describes how finely the spectra were sampled, not what acquired them, and a
dense profile spectrum can come from a MALDI-TOF, a TOF-SIMS, or an Orbitrap
run in profile mode. SCiLS Lab does not guess modality either -- its importer
takes it as an argument (--project TIMSTOF|TOF|FT).
The timsTOF flag is a case-insensitive substring match on the instrument name,
because Bruker names the family many ways (timsTOF fleX MALDI-2,
timsTOF Pro 2, timsTOF SCP, and so on).
Which detector wins¶
Detectors are tried in a fixed priority order and the first match wins: timsTOF, Rapiflex, FT-ICR, Orbitrap, PHI, generic centroid, then a catch-all default. This table is the actual observed behaviour of that chain:
| Metadata | Detector | Method | Axis type |
|---|---|---|---|
| timsTOF, centroid | timsTOF | nearest_neighbor |
reflector_tof |
| timsTOF, profile high-density | timsTOF | nearest_neighbor |
reflector_tof |
| Rapiflex, profile | Rapiflex MALDI-TOF | tic_preserving |
constant |
| Bruker MALDI-TOF | Rapiflex MALDI-TOF | tic_preserving |
constant |
PHI SmartSoft-TOF .raw |
PHI SmartSoft-TOF (ToF-SIMS) | nearest_neighbor |
linear_tof |
| unknown vendor, profile (any density) | Unknown (default) | nearest_neighbor |
constant |
| unknown, centroid | ImzML Centroid | nearest_neighbor |
reflector_tof |
| no usable metadata | Unknown (default) | nearest_neighbor |
constant |
Why PHI needs its own row
Without it PHI reaches the catch-all default and is reported as
constant. Thyra's own answer would still be nearest_neighbor, so
nothing breaks inside Thyra -- but preview_msi hands that axis type
to callers, and a caller that maps constant to profile-MALDI
conventions arrives at tic_preserving onto an equidistant axis. That
combination is destructive here: a PHI pixel holds a median of 44
measured points across m/z 0.5--1850, and interpolating between them
fabricates intensity in every bin in the gaps, with the TIC rescale
hiding it behind a total that still balances. The detector reports the
law the data actually follows -- PhiMassAxis steps at a constant
flight time, so spacing goes as sqrt(m/z) -- which is linear_tof.
tic_preserving is gated on the source and target axis laws matching
A detector may only ask for tic_preserving if it knows the spacing law
of the grid the spectra arrive on, and that law is the one it is asking
the target axis to use. Only the Rapiflex route qualifies:
RapiflexReader lays every spectrum out with np.linspace, so the source
is uniform in m/z, and the axis it requests is constant -- the same law.
Anything else is refused and gets nearest_neighbor instead, with a log
line saying so. This is the rule SCiLS Lab applies: TIC-preserving
resampling "if all axis types are identical", otherwise interpolation
(2026b User Guide, p.80). It is also exactly when Thyra's
interpolate-then-rescale operator is exact -- see the !!! danger box
below for what it costs off the diagonal.
The Orbitrap and FT-ICR detectors cannot currently fire
Both match on an instrument_type of "Orbitrap" or "FT-ICR", but no
reader populates that key: the only place it is set is the Rapiflex reader,
which hardcodes "MALDI-TOF". Neither the imzML nor the Bruker .d
metadata extractor emits it at all.
Still true: reading the declared MS:1000127/MS:1000128 accession fixed
how spectrum representation is detected, not how instrument_type is
populated. Nothing writes "Orbitrap" or "FT-ICR" into it, so these two
detectors remain unreachable. What did change is which row such data lands
on: a profile-mode Orbitrap file now correctly reports profile, so it falls
to the default row (constant) rather than being misread as centroid and
given reflector_tof.
So Orbitrap and FT-ICR data does not get the orbitrap or fticr
axis automatically -- it falls through to the generic centroid or default
row above and gets reflector_tof or constant. Set
--mass-axis-type orbitrap or --mass-axis-type fticr explicitly if you
want the matching physics.
tic_preserving is selected for profile MALDI-TOF, not for high resolution
It is easy to assume the high-resolution analysers get the more elaborate
method. They do not. tic_preserving is chosen for profile data, where
a peak is spread over many points and rebinning would otherwise change the
total ion count. Orbitrap and FT-ICR data is normally centroided, so it gets
nearest_neighbor, which is the correct choice for discrete peaks. If you
have profile Orbitrap or FT-ICR data, set
--resample-method tic_preserving yourself.
Do not combine tic_preserving with a non-uniform axis type
tic_preserving interpolates onto the target axis and then applies a
single scaling factor to the whole spectrum. A single factor cannot
account for bin widths that vary across the mass range, so pairing it with
linear_tof, reflector_tof, orbitrap or fticr suppresses high-m/z
ions relative to low-m/z ones. Measured across 300-1100 m/z, two ions of
equal abundance come back with their ratio distorted by 1.9x on
linear_tof, 3.7x on reflector_tof, 7.0x on orbitrap and 13.4x on
fticr.
Auto-selection cannot produce these pairings. tic_preserving is only
ever chosen alongside constant, and only on a source grid that is
itself uniform in m/z, which is what makes it exact; the detector chain
enforces that. You have to ask for the combination with two explicit
flags, and Thyra takes you at your word.
If you want a non-uniform axis, use nearest_neighbor, which moves each
peak into a single bin and is unaffected by bin width.
The chosen detector, method, and axis type are all logged:
INFO - Detected instrument type: timsTOF
INFO - Selected resampling method: NEAREST_NEIGHBOR
INFO - Selected axis type: REFLECTOR_TOF
Methods¶
nearest_neighbor¶
Each target bin takes the intensity of the nearest original m/z value. Peaks stay sharp and stay put, and no intensity is invented between them. This is the right choice for centroid data, where a peak is a single discrete mass and smearing it across neighbouring bins would be wrong.
Consequence worth knowing: on a target axis finer than the source spacing, most bins receive nothing. A dataset resampled from 4,000 source points onto 190,000 bins has exactly 4,000 populated bins per spectrum, and the rest are zero. This is why you should extract ion images by summing over an m/z window rather than picking the single nearest bin -- see the tutorial.
tic_preserving¶
Linear interpolation onto the target axis, followed by rescaling so the spectrum's total ion current matches the original. This is the right choice for profile data: a peak spans many points, interpolation reconstructs its shape on the new grid, and the rescaling step stops rebinning from quietly changing quantitation.
Use it whenever the total ion count per pixel has to stay comparable before and after conversion.
The total that is preserved is the share of the spectrum inside the axis range.
If you crop with --resample-min-mz or --resample-max-mz, intensity outside
the window is dropped rather than redistributed over the bins you kept, so the
per-pixel total falls by whatever you cropped out -- a cropped window should not
claim ions from the parts that were cut away. That share is measured by area, so
a window narrower than the spacing between source points still keeps a
proportionate amount instead of collapsing to zero.
With the default axis, which spans the dataset's own mass range, nothing is dropped and the total is preserved exactly.
Gaps in the source m/z values¶
Interpolation has no notion of a gap. Between the last source point before an
empty stretch and the first one after it, np.interp draws a straight line,
and every target bin in between gets an intensity that was never measured. On
sparse or thresholded m/z arrays that is most of the axis: forcing
tic_preserving onto bellini.imzML -- 2,222 points per spectrum, median
source spacing 0.0072 Da, largest gap 1,269 Da -- puts 80.6% of the output
total ion current more than 0.5 Da from any measured point.
--resample-gap-tolerance sets how far a target bin may sit from the nearest
source m/z before its interpolated value is discarded instead of trusted. It is
the same parameter Cardinal calls tolerance and matter calls tol.
Masking happens before the TIC rescale, so the intensity returns to the bins that do have measurements behind them rather than being deleted; the per-pixel total is still preserved.
Sizing it: on a source grid of uniform step s, no target bin is ever more
than s/2 from a source point, so any tolerance above half the widest gap in
your source m/z values changes nothing. Continuous profile data is dense and
evenly sampled, which is why the default is no limit -- there is nothing to
refuse. Set it when your m/z arrays are sparse or peak-picked and you are
asking for tic_preserving anyway.
nearest_neighbor needs no such parameter: it only ever fills a bin some peak
was snapped into.
Axis types¶
The axis type sets how bin width grows with m/z. Each corresponds to the physics of a mass analyser, so that bins track the instrument's real resolving power instead of over-sampling the low end and under-sampling the high end.
| Axis type | In SCiLS Lab | Bin width scales as | Rationale |
|---|---|---|---|
constant |
Constant (equidistant) | constant Da | Equidistant bins |
linear_tof |
Axial TOF | sqrt(m/z) |
Linear TOF: flight time t ∝ sqrt(m/z), so equal time bins give sqrt(m/z) mass bins |
reflector_tof |
Orthogonal TOF | m/z |
Constant relative resolution R = m/Δm; bins are uniform in ln(m/z) |
orbitrap |
Orbitrap | m/z^1.5 |
Orbitrap frequency f ∝ 1/sqrt(m/z), so equal frequency bins give m/z^1.5 mass bins |
fticr |
MRMS (Fourier-transform) | m/z^2 |
Cyclotron frequency f ∝ 1/(m/z), so equal frequency bins give m/z^2 mass bins |
The five types, and their scaling laws, are the ones SCiLS Lab offers for the
common mass axis (2026b User Guide, p.75). Thyra's names are SCiLS's older
ones: linear_tof and reflector_tof were "Linear TOF" and "Reflector TOF"
in earlier SCiLS versions and are now Axial TOF and Orthogonal TOF, and the
FT-ICR type is now called MRMS. The laws are unchanged; only the labels moved.
reflector_tof is the most broadly useful of these: constant relative
resolution means constant relative mass accuracy across the whole range, which
is what most MS workflows assume.
Bin count¶
You can set the bin count directly, or specify a target bin width at a reference m/z and let Thyra derive the count from the axis physics. The two are mutually exclusive.
When you specify neither, these defaults apply:
| Axis type | Default width | At reference m/z |
|---|---|---|
linear_tof |
17 mDa | 300 |
| everything else | 5 mDa | 1000 |
The 17 mDa / m/z 300 pairing for linear_tof was chosen to be close to the
axis SCiLS Lab produces for FlexImaging data. Treat it as a working default
rather than a reproduction: the figure does not appear in the SCiLS Lab
2026b User Guide, and no version that states it has been identified, so
nothing here is a claim of exact equivalence. The code comment implementing it
says "SCiLS-like ~17 mDa", which is the accurate description. If you need to
match a particular SCiLS project, set --resample-width-at-mz and
--resample-reference-mz from that project's own axis.
The count is then derived per axis type:
| Axis type | Bin count |
|---|---|
reflector_tof |
ln(max/min) * (ref_mz / width) |
linear_tof |
(2/k) * (sqrt(max) - sqrt(min)), k = width / sqrt(ref_mz) |
orbitrap |
2 * (1/sqrt(min) - 1/sqrt(max)) * (ref_mz^1.5 / width) |
fticr |
(1/min - 1/max) * (ref_mz^2 / width) |
constant |
(max - min) / width |
Each is the integral of 1 / width(m) across the mass range, so the bin width
the axis actually realizes at your reference m/z is the width you asked for.
The result is floored at 100 bins.
A worked example, from the tutorial's synthetic dataset --
250-1200 Da, constant axis, default 5 mDa width:
which is what the log reports:
INFO - Calculating bins for 5.0 mDa width at m/z 1000.0
INFO - Calculated 190000 bins for constant axis type
INFO - Building resampled mass axis: 250.00 - 1200.00 m/z, 190000 bins
190,000 bins from 4,000 source points sounds extravagant, and on the
nearest_neighbor path -- which is what auto picks for this dataset -- it
very nearly is free. Each source point lands in exactly one bin, so the
non-zero count does not depend on the bin width at all and only the index
arrays lengthen: the tutorial store comes out at roughly 31 MB, and going from
4,000 bins to 190,000 costs about +0.8% on disk.
That only holds for nearest_neighbor
tic_preserving interpolates, so it populates essentially every bin. The
matrix comes out 99.998% dense and the store becomes
n_pixels x target_bins x ~7.5 bytes -- independent of how sparse the
source was. Measured on this same 4,000-point source: at 190,000 bins it
costs 1.44 MB per pixel, so 400 pixels take 574 MB against 6.93 MB at
4,000 bins, an 82.9x blowup, and the 1,728-pixel tutorial dataset would
take roughly 2.5 GB.
On that path the default bin count is a decision rather than a detail.
Pass --resample-bins or --no-resample if you would rather not upsample.
The 10 million bin cap¶
--no-resample on a processed-mode imzML does not build a physics axis at all.
It builds the raw axis: every distinct m/z value in the whole dataset, one
column each. When the peak lists genuinely share m/z values that is small. When
they do not -- which is the normal case for centroided peak picking, where two
pixels almost never report a peak at the same double -- it grows to roughly one
column per peak in the entire dataset, and there is no natural stopping point.
Thyra gives up once that axis passes 10 million unique m/z values, and says what to do instead:
ValueError: Common mass axis exceeded 10,000,000 unique m/z values after
12,400 of 918,855 spectra (10,004,112 so far). The peak lists in this dataset
do not share m/z values, so a raw axis grows to roughly one column per peak,
which is not usable downstream. Convert with resampling instead (it is the
default; --no-resample disables it), or raise max_mass_axis_length.
10 million is SCiLS Lab's own limit on the same quantity: "Data sets in SCiLS Lab are limited to a maximum of 10 million bins on the common mass axis" (2026b User Guide, p.76).
The cap is Python-API only, through reader_options, and applies to the imzML
reader:
from thyra import convert_msi
convert_msi(
"data.imzML",
"out.zarr",
reader_options={"max_mass_axis_length": 50_000_000}, # or None for no cap
)
It has no effect when resampling is on, which is the default: a resampled axis has the bin count you asked for.
Overriding the automatic choice¶
Every part is independently overridable; anything you leave alone stays automatic.
# Method and axis type. nearest_neighbor is the pairing to use with a
# non-uniform axis -- see the warning under "Selection" above.
thyra input.imzML output.zarr \
--resample-method nearest_neighbor \
--mass-axis-type orbitrap
# Fixed bin count
thyra input.imzML output.zarr --resample-bins 50000
# Target width at a reference m/z instead of a bin count
thyra input.imzML output.zarr \
--resample-width-at-mz 0.01 \
--resample-reference-mz 500
# Restrict the mass range
thyra input.imzML output.zarr \
--resample-min-mz 400 --resample-max-mz 1000
# Off entirely
thyra input.imzML output.zarr --no-resample
The Python API does not resample by default
convert_msi() called without resampling_config keeps the original mass
axis, whereas the thyra command-line tool resamples by default. Pass a
resampling_config explicitly if you want the CLI's behaviour from Python.
Checking what was used¶
The decision is recorded in the output store, so a converted dataset is self-describing:
import spatialdata as sd
sdata = sd.read_zarr("output.zarr")
table = sdata.tables["msi_dataset_z0"]
mz = table.var["mz"].values
print(f"{len(mz)} bins, {mz.min():.2f} -- {mz.max():.2f} m/z")
# Bin width across the range tells you the axis type in practice:
# flat -> constant, growing linearly -> reflector_tof, and so on.
import numpy as np
d = np.diff(mz)
print(f"bin width: {d.min()*1000:.3f} -- {d.max()*1000:.3f} mDa")
For the tutorial dataset this prints a flat 5.000 mDa across the range,
confirming a constant axis at the default width.
Practical guidance¶
- Leave it alone unless you have a reason. The detected settings are right for the common Bruker and imzML cases.
- Disable it (
--no-resample) when you want the untouched vendor axis -- your own peak picking, centroiding, or calibration downstream. - Set
tic_preservingif quantitation matters and your data is profile, especially profile data from a high-resolution analyser, which the automatic choice will not pick it for. - Set
--resample-binsany time you need to match another tool's axis exactly. - Narrow the mass range (
--resample-min-mz/--resample-max-mz) before increasing bin counts; it is usually the cheaper way to get the resolution you need where you need it.
See also¶
- CLI Reference -- the flags
- Tutorial -- resampling in the context of a full conversion
- Output Format -- where the m/z axis lives in the store
- API Reference --