Thyra reads Bruker solariX / MRMS (FT-ICR) imaging .d directories natively by consuming peaks.sqlite, the processed peak store ftmsControl writes into every imaging acquisition. There is no public specification for this file; the layout below was established by inspecting real acquisitions spanning 2018-2024 (ftmsControl 2.2, flexImaging 5.0.80-5.0.89) and decode-verifying the binary blobs against their own metadata. The same file is read by at least one independent open-source tool for the solariX XR, so the store is not a single-instrument quirk.

What this route deliberately is not: a transient processor. The .d also holds the raw time-domain transients (ser, tens of GB per region) and Bruker's proprietary .mcf containers. Thyra never opens either -- no FT, no apodization, no peak picking. The peaks were picked by the acquisition software at acquisition time, and that processing is frozen in the data. Anyone needing different peak picking must export via the vendor software.


Directory layout

sample.d/
  peaks.sqlite            centroided per-pixel peak lists  <- what Thyra reads
  ImagingInfo.xml         per-scan index (spot name, TIC, base peak)
  ser                     raw transients, fixed-length per scan (ignored)
  <guid>_N.mcf/.mcf_idx   Bruker container stores (ignored)
  parameterChanges.sqlite per-scan parameter history (ignored)
  Storage.mcf_idx         container index (ignored)
  *.m/                    acquisition method directory (ignored)
sample.mis                flexImaging sequence NEXT TO the .d, same stem
sample.dat                flexImaging cache (ignored)

Detection requires peaks.sqlite and ImagingInfo.xml inside a .d directory. Two traps are handled explicitly:

  • Pre-scan directories. flexImaging writes sibling .d directories such as sample_0_C17_000001.d holding fid + analysis.baf + calib.bin -- single-spectrum pre-scans, not imaging runs. They contain none of the detection files and are rejected as unrecognised Bruker data.
  • Peaks-less acquisitions. A .d with ser and ImagingInfo.xml but no peaks.sqlite is still recognisably solariX-family; the error names the imzML export fallback (DataAnalysis, SCiLS Lab, or flexImaging) instead of claiming the directory is not Bruker data.

peaks.sqlite

Three tables matter. Properties is a key/value table carrying the schema identity (SchemaType = Imaging, SchemaVersionMajor = 1, SchemaVersionMinor = 2 in every file inspected), the acquisition software name/version, the instrument vendor, family, source type and serial number, the operator, the acquisition timestamp and the acquired m/z range (MzAcqRangeLower/MzAcqRangeUpper).

The reader checks SchemaType and SchemaVersionMajor at open and refuses anything it has not been verified against, quoting the values it found. A newer ftmsControl that bumps the major version fails loudly rather than decoding blobs on an unknown layout.

AcquisitionKeys holds one row per acquisition mode with Polarity, ScanMode, AcquisitionMode and MsLevel. Only the polarity enum has been verified (against a paired positive/negative acquisition of the same sample): 0 = positive, 1 = negative. The other enums are undocumented, so Thyra stores the raw integers without inventing labels. Observed values: ScanMode 0 / MsLevel 1 on full scans, ScanMode 2 / MsLevel 2 on a CASI (quadrupole isolation) run.

Spectra holds one row per pixel: stage-raster coordinates (XIndexPos/YIndexPos), NumPeaks, per-scan acquisition values (NumSummations, LaserPower, LaserRepRate), RegionNumber, and the peak blobs.

Blob encoding

Column Encoding Length
PeakMzValues little-endian float64, ascending NumPeaks * 8
PeakIntensityValues little-endian float32 NumPeaks * 4
PeakFwhmValues little-endian float32 NumPeaks * 4
PeakSnrValues little-endian float32 NumPeaks * 4

The m/z values are already calibrated -- the frequency-to-m/z law (ML1/ML2/ ML3 in the acquisition method) has been applied by the time the peaks are stored, so the reader does no calibration math. Every blob length is checked against NumPeaks * itemsize before decoding; a mismatch is refused as corruption rather than silently truncated. PeakFlags, BH and BC are undecoded (flags were all-zero in every file inspected) and nothing is derived from them.

The database is opened strictly read-only through a URI (mode=ro), so sqlite never creates journal or WAL side files next to it. That matters because solariX data commonly lives on read-only network shares. One consequence of that: a mapped network drive resolves to a UNC path (\\server\share\...) before the reader sees it, and in a file: URI the server name would read as an authority, which sqlite rejects. The reader therefore writes such paths as file:////server/share/... -- empty authority, path intact -- which opens where file://server/share/... does not.

Coordinates and sparsity

XIndexPos/YIndexPos are absolute stage-raster indices -- a real region started at X 1239, not 0. The reader subtracts the per-dataset minimum to produce 0-based coordinates and reports the subtracted origin in coordinate_offsets and the absolute extents in the format metadata.

Rasters are sparse: flexImaging acquires the polygon drawn on the optical image, so the bounding box contains cells that were never measured (81.5 % fill on the reference region). Unmeasured cells are simply absent from the output, exactly like sparse pixels in any other format. A measured scan can also record zero peaks (seen at the tissue-free edges of one acquisition); such pixels are skipped like unmeasured ones.

One fidelity check worth recording: on the reference region the sum of a pixel's stored peak intensities reproduces the per-scan tic value in ImagingInfo.xml -- an independent record written at acquisition time -- to a ratio of 0.9999. TIC images computed from the peak store are therefore faithful to what the instrument reported.

Pixel size

The .d itself does not record the raster pitch anywhere -- the acquisition method file was swept for raster/pixel tokens and carries none. The only source is the <Raster>x,y</Raster> element (micrometres) of the flexImaging .mis file sitting next to the .d with the same stem.

Resolution order: a stem-matching .mis in the parent directory; failing that, a lone .mis in the parent is accepted as unambiguous. Multiple non-matching candidates are refused -- on a multi-region slide each region has its own .mis, and guessing one would silently assign a wrong pixel size. When no source resolves, the pixel size is reported as unknown and conversion asks for --pixel-size instead of defaulting.

Resampling

The stored spectra are centroided per-pixel peak lists, so every spectrum has its own m/z values. The metadata extractor stamps instrument_type = "FT-ICR" from the instrument identity in Properties (InstrumentFamily 513 additionally maps to the model name solariX; unknown family numbers are surfaced raw). The resampling decision tree therefore lands on nearest_neighbor onto the fticr axis with no flags -- the same pair a vendor imzML export of the same data reaches, minus the export step. See Resampling for why TIC-preserving is refused for centroid data.

Cross-validation against a vendor imzML export

The reader was validated against a SCiLS-lineage centroid imzML export of the same acquisition (a 7,362-pixel run; the export is a 949-pixel tissue ROI crop with re-based coordinates). Registering the ROI onto the reader's raster by normalized cross-correlation of TIC images -- testing all four axis orientations -- found the identity orientation with a pure translation, and every ROI pixel landed on a measured pixel. On that overlap:

  • The peak lists are the same peaks. Every one of the 949 overlapping pixels has an identical peak count on both sides -- the exporter consumed the same picked peaks this reader decodes, it did not re-pick.
  • Intensities differ only by the exporter's normalization. Within each pixel the imzML/native intensity ratio is constant to the last float32 digit, and the per-pixel factor times the RMS of the pixel's native peak intensities is one dataset-wide constant: the export applied RMS normalization (the _rms in its filename) plus a global scale. Dividing the factor out reproduces the native TICs with a maximum relative error of 2e-7, and the TIC correlation goes to r = 1.000000.
  • m/z values agree to 0.00 ppm on average. The signed difference is zero-mean in every 200 Da band -- no calibration difference. Individual peaks deviate by up to ~4 mDa, growing quadratically with m/z (0.06 mDa median at m/z 200, 2.0 mDa at m/z 1200), which is the FT-ICR grid-spacing law: the exporter snapped centroids to its own profile-grid axis. The native reader keeps the full-precision centroids as stored.

So the native route and the vendor-export route see the same data; the export adds RMS normalization and axis snapping, and drops the pixels outside the drawn ROI.

What the native route adds over an imzML export

Everything an export carries, plus metadata no exporter writes: per-pixel laser power, repetition rate and summation counts, the acquisition software identity and version, the raw acquisition-mode enums, per-scan region numbers, the absolute stage-raster extents, and the acquisition m/z range as acquired. And no manual vendor-software step per dataset.