Change how Thyra converts

Thyra chooses its settings from your data, and most datasets need no options at all. This page explains the options you are most likely to want, one task at a time.

Add an option to the end of the command, after the result's name:

thyra my_slide.d my_slide.zarr --no-optical

Set the pixel size

Thyra reads the pixel size from your files. If they do not record it, the conversion stops before it writes anything. Give the size in micrometres:

thyra input.imzML output.zarr --pixel-size 20

The same size is used for both directions.

Advanced: pixels that are not square

A pixel size read from the file can differ between x and y, and Thyra keeps both. --pixel-size always gives square pixels.

Convert one region

A Bruker timsTOF slide with several regions is converted as one result, with every pixel labelled with its region. To convert only one region, add --region and the region's name as flexImaging shows it:

thyra slide.d region_03.zarr --region 03

Thyra lists the regions it finds when it starts, so a first run shows you the names. If the name matches none of them, Thyra stops and lists the names it did find.

Advanced: region numbers, and the cropped photo

If the value matches no region name, Thyra reads it as the region's number in the Bruker database, which starts at 0. The names and the numbers often differ: region 03 need not be number 3. When you convert one region, Thyra also crops the slide photo to that region, plus a 10% margin.

Other formats refuse --region. From Python, thyra.list_regions lists the regions of any input without converting it.

Leave out the optical image

For Bruker data, Thyra adds every image it finds in the data folder and the folder above it. The image named in the flexImaging .mis file is the one the MSI data is lined up with. To leave all images out, add --no-optical.

Choose the m/z axis

What Thyra does by default

Each spectrum is recorded at slightly different m/z values. To compare pixels, Thyra puts every spectrum on one shared set of m/z positions, called bins. Each column of the result then means the same m/z in every pixel. Most analysis tools need this. The step is called resampling.

Thyra chooses the width of the bins, and the way peaks are moved onto them, from your instrument. For most instruments the bins are 5 mDa wide at m/z 1000.

Advanced: what Thyra chooses for each instrument
Your data Method Axis type Bin width at m/z 1000
Bruker timsTOF nearest_neighbor reflector_tof 5 mDa
Bruker rapifleX tic_preserving constant 5 mDa
Bruker solariX, or imzML from an FT-ICR nearest_neighbor fticr 5 mDa
imzML from an Orbitrap nearest_neighbor orbitrap 5 mDa
PHI ToF-SIMS nearest_neighbor tof about 57 mDa, 3 bins per peak
Waters SELECT SERIES MRT (profile, the default) tic_preserving linear_tof 1.3 mDa
Waters SELECT SERIES MRT, centroid nearest_neighbor tof 1.75 mDa
Other Waters (centroid, the default) nearest_neighbor reflector_tof 2 mDa
Anything else, centroid nearest_neighbor reflector_tof 5 mDa
Anything else, profile nearest_neighbor constant 5 mDa

The axis type sets how the bin width changes across the mass range. Resampling explains each type and each method.

Keep the original m/z values

Add --no-resample. Each spectrum then keeps its own m/z values. Use it when you want to do your own peak picking or calibration.

Make the bins finer or coarser

Set the bin width at m/z 1000, in Da. A smaller width gives finer bins:

thyra input.imzML output.zarr --resample-width-at-mz 0.002

Finer bins make a larger result and need more memory. If an axis would need more memory than your computer has free, Thyra refuses it before the conversion starts.

Advanced: set the number of bins instead

--resample-bins 100000 sets the number of bins directly. Use one of the two options, not both. --resample-reference-mz moves the m/z at which --resample-width-at-mz applies (1000 by default).

Keep only part of the mass range

thyra input.imzML output.zarr --resample-min-mz 300 --resample-max-mz 1200

Both limits are included. Peaks outside the range are left out, and Thyra warns you once.

Advanced: choose the method and the axis type yourself

--resample-method chooses how peaks are moved onto the bins:

  • nearest_neighbor puts each peak in the nearest bin. It is safe on every axis type.
  • tic_preserving spreads the signal between bins and keeps the total ion current. It is exact only on an axis spaced like your instrument's own sampling. Bins wider than that sampling still keep narrow peaks.

--mass-axis-type chooses how the bin width changes across the mass range. If you change it and leave the method on automatic, Thyra switches to nearest_neighbor where it has to. If you ask for tic_preserving yourself, Thyra does what you ask. Read Resampling before combining the two options.

Keep 3D data together

A dataset with several slices becomes one result part per slice, named _z0, _z1 and so on. To keep the slices together as one 3D volume, add --handle-3d. Give the distance between slices, in micrometres, with --z-spacing:

thyra volume.imzML volume.zarr --handle-3d --z-spacing 20

Without --z-spacing, Thyra assumes the slices are one pixel apart, and warns you.

Name the dataset

Every part of the result is named after the dataset, msi_dataset unless you choose another name:

thyra slide.d slide.zarr --dataset-id mouse_brain

Use letters, digits, _, . and -.

Keep a record of the conversion

To save everything Thyra reports into a file, add --log-file:

thyra slide.d slide.zarr --log-file conversion.log

Add -v DEBUG for more detail. The file is added to, not replaced, on every run.

Advanced: exit status, for scripts

thyra exits with status 0 when the result was written, and 1 when the conversion failed or was refused. It exits with 2 when the command itself is wrong, for example an option that does not exist or a result that already exists.

Options for one kind of instrument

Advanced: options for imzML, Bruker and Waters data
  • --spectrum-type profile or centroid (imzML): use it when the file declares the wrong type of spectrum. It changes which m/z axis Thyra chooses.
  • --no-recalibrated (Bruker timsTOF): use the calibration recorded during the acquisition, instead of the newest recalibration.
  • --tdf-spectrum vendor_centroid (Bruker timsTOF with ion mobility): use Bruker's own peak-picked spectrum instead of the sum of all scans.
  • --mobility-grid (Bruker timsTOF with ion mobility, or processed imzML with a mobility array): add a table that keeps the ion mobility of every pixel.
  • --no-msms-table (Bruker timsTOF PASEF MS/MS): leave out the table that splits the fragments by precursor.
  • --waters-spectrum profile or centroid (Waters): profile keeps close masses apart, at about three times the size. It is the default on a SELECT SERIES MRT; other Waters instruments default to centroid.
  • --intensity-threshold (any data): leave out peaks below this intensity. They cannot be recovered from the result.

Convert from Python

Advanced: the same settings in Python

convert_msi takes the same settings as the command:

from thyra import convert_msi

convert_msi(
    "slide.d",
    "slide.zarr",
    pixel_size_um=20,
    resampling_config={},
)

From Python, resampling is off unless you pass resampling_config. An empty {} gives the same automatic choice as the command. The Python API page lists every argument. convert_msi returns True when the result was written and False when it was not.